SENSOFIT INS 1200
Insertion assembly with pressure indication and integrated safety features for analytical sensors
- For use with Ø12 / 120 mm or Ø0.47 / 4.7" sensors (pH, ORP etc.)
- Integrated pressure test via "Quick'n'Safe" mechanism for safe, fast and comfortable removal and installation of sensors; enables sensor exchange without risk of glass breakage in <15 s
- Sensor connection: PG13.5; Process connection: ¾ or 1 NPT, G¾
- Max. +95°C / +203°F and 12 bar / 174 psi
Overview
The SENSOFIT INS 1200 is a patented insertion assembly for 12 mm / 0.47" analytical sensors. It is particularly suitable for use in pressurised applications with increased process and occupational safety demands. Its "Quick'n'Safe" mechanism makes SENSOFIT INS 1200 the first assembly for direct mounting that detects and indicates current line pressure. It also features a built-in pressure check before a sensor is removed for maintenance.
Product features and options depend on device configuration: Contact us to make sure that your preferred functionality can be combined in one device.
- Cost-effective direct mounting system for safe, quick and comfortable installation and removal of 12 mm / 0.47" analytical sensors
- Integrated safety
- Pressure indication with warning sign – prevents the user from removing the sensor from a pressurised line
- Test for remaining process pressure before each sensor removal
- High degree of occupational safety due to two-stage removal movement with safe intermediate position (Quick'n'Safe system)
- Enables controlled draining of medium at low residual pressure
- No fluid leakage in user direction in case of damaged seal
- High operating comfort
- Quick installation and removal due to short 60° turning movement
- No twisting and torsion of cable, no wear on plug contacts
- Significant time savings on installation and maintenance work
- Sealing of the connection thread is carried out only once during initial installation; no resealing necessary after sensor exchange
- High tightening torque for permanent tightness
- Sensor exchange without risk of glass breakage in less than 15 seconds with second sensor / mounting set, e.g. pre-calibrated SMARTPAT pH and ORP sensors
- Mountable to ¾ or 1 NPT thread (or G¾), e.g. as a direct replacement for a sensor with plastic housing and male thread
- Individually configurable: from robust, cost-effective stainless steel version to corrosion- and solvent-resistant design with wetted components made of PPS/PVDF and FFKM O-rings